Subject: X-RAY TUBE INSERT PROTECTION CURVES tech. Пожалуйста, помогите перевести.X-RAY TUBE INSERT PROTECTION CURVES Выражение встречается в следующем контексте: SECTION 3 X-RAY TUBE SELECTION 3.1 X-RAY TUBE INSERT PROTECTION CURVES In order to properly select the X-ray Tube Insert Protection Curves for the Tubes connected to the Generator, perform the following procedure: 1. Select one workstation of the corresponding X-rayTube to be configured. 2. Enter in calibrationmodeand select the respectivememory location (E02 or E18) (memory location is shown on the mA Display). 3. Identify in Section 4 “X-ray Tube Data”, the X-ray Tube that is being installed and note its Tube type number. 4. Set the Tube number by pressing the “Increase” or “Decrease” Density push-buttons until the correct number is showed in the kV Display. 5. Store the value by pressing the “AEC Reset” push-button. 6. Verify that the Tube code (ID) showed in the mAs Display is the same of the Tube code listed in Section 4 “X-ray Tube Data”. The Tube code (ID) can be only read for the selected X-ray Tube after pressing the “AEC Reset” push-button. 7. If required, repeat this procedure for the other X-ray Tube. 8. Exit calibration mode. Record configuration data for E02 and E18 in the Data Book. Заранее спасибо
|